Home>IEC standards>BS EN IEC 61326-2-2:2021 pdf free

BS EN IEC 61326-2-2:2021 pdf free

BS EN IEC 61326-2-2:2021 pdf free.Electrical equipment for measurement, control and laboratory use – EMC requirements Part 2-2: Particular requirements一Test configurations, operational conditions and performance criteria for portable testing, measuring and monitoring equipment used in low-voltage distribution systems.
Examples of such EUT5 include, but are not limited to, voltage detectors, insulation testers, earth continuity testers, earth resistance testers, leakage current clamps, loop impedance testers, ‘residual-current-device-testers” (RCD-testers) and phase sequence testers as defined in IEC 61557 (all parts).
NOTE Particular EUC requirements for equipment covered by IEC 61557-8 and IEC 61557-9 are given in IEC 61326-2-4,
The manufacturer specifies the environment for which the product is intended to be used and/or selects the appropriate test level specifications of IEC 61326-1.
2 Normative references
The following documents are referred to in the text in such a way that some or all of their content constitutes requirements of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.
Clause 2 of IEC 61326-1:2020 applies, except as follows:
Addition:
IEC 61 326-1:2020, Electrical equipment for measurement, control and laboratory use — EMC requirements — Part 1: General requirements
IEC 61557 (all parts). Electrical safety in low voltage distribution systems up to 1 000 V AC and 1 500 V DC — Equipment for testing, measuring or monitoring of protective measures
3 Terms and definitions
For the purposes of this document, the terms and definitions given in lEG 61326-1:2020 and IEC 60050-16 1 apply.
4 General
Clause 4 of IEC 61 326-1 :2020 applies.
5 EMC test plan
5.1 General
Subclause 5.1 of IEC 61326-1:2020 applies.
5.2 Configuration of EUT during testing
Subclause 5.2 of IEC 61326-1:2020 applies, except as follows:
Addition:
5.2.4.101 Test and measurement 110 ports
Electrostatic discharge shall be applied to the mated connector or the shield of the unmated port, but not to the inner pins of shielded port or cable connectors (for example. BNC, Dsubminiature, GPIB, RS232, USB, etc.).
For the test according to IEC 61000-4-3, the following conditions shall be met. Test and measurement ports shall be connected with test leads recommended or supplied with the EUT. Where the test leads are unspecified, typical test leads shall be used. The test leads shall be connected and arranged in a typical configuration for each operation mode, according to Figure 101.
If the test leads recommended or supplied are longer than I m, each one should be bundled so that the test object is in a horizontal distance of 1 m to the ELJT.
The test leads shall be arranged 0,1 m apart in a horizontal position on the test table.BS EN IEC 61326-2-2 pdf free download.